Aerogen
®
Solo System Instruction Manual
9
Control / Indicator Function
30 Min. Indicator
• Green (steadily lit) = 30 Minute
nebulisation cycle on
• Green (ashing) = Low battery
power
• Nebuliser automatically powers off after
30 minutes have elapsed
Continuous Indicator
• Green (steadily lit) = Continuous
nebulisation cycle on
• Nebuliser does not power off
automatically
Error Indicator
• Amber (steadily lit) = Aerogen Solo
nebuliser disconnected from Aerogen
Pro-X Controller
• Amber (flashing) = Aerogen Pro-X drive
voltage error
On/Off Power Button
• To operate in 30 Minute Mode, press the
On/Off button once
• To operate in Continuous Mode, press
and hold the On/Off button for greater
than 3 seconds from off
• Pressing during nebulisation turns off
power to the nebuliser
Battery Status Indicator
• Green = Battery fully charged
• Amber = Battery charging
• No light = Battery in operation
Recharging the Battery
To recharge the battery, connect the AC/DC adapter to the controller and
connect to AC power source. The battery status indicator is amber while
charging and green when fully charged.
If the controller is placed in long-term storage, it is recommended that the
battery be recharged every 3 months.
Allow a minimum of four hours for the internal battery to fully recharge.