Troubleshooting
If these suggestions do not correct the problem, discontinue use of any
device and contact your local Aerogen sales representative.
Table 2. Aerogen Pro-X Controller Troubleshooting
If this happens: It could mean: Try this:
The 30 Min. indicator
flashes during
nebulisation.
Battery power is low.
Recharge battery (see
Recharging the Battery).
Battery will not
recharge.
Controller is connected
to the AC/DC adapter
and the battery charging
light is illuminated green
and the 30 Min. indicator
light is flashing.
It may be time to replace
the battery.
Contact your local Aerogen
sales representative.
Battery will not retain
initial charge.
Rechargeable battery may
need to be replaced.
Contact your local Aerogen
sales representative.
The 30 Min. or
Continuous light
illuminates, but aerosol
is not visible.
No medication in nebuliser.
Refill medication through filler
cap in the nebuliser (see page
17).
It may be time to replace
the nebuliser.
See Warranty and Life of
Product. Refer to Aerogen
Solo parts list by visiting
www.aerogen.com.
30 Min. or Continuous
indicator does not light
when On/Off power
button is pressed.
There is no power to the
system.
Verify that AC/DC adapter
is securely attached to
controller.
Rechargeable battery is
depleted.
Recharge battery (see
Recharging the Battery).
The fault indicator light
illuminates.
The controller cable is
incorrectly connected to
the nebuliser, or electronics
are malfunctioning.
Verify that controller cable
is correctly connected to
both the nebuliser and the
controller.