Aerogen
®
26
Specifications
Table 3. Physical Specifications of the Aerogen Solo System
Nebuliser Dimensions
67 mm H x 48 mm W x 25 mm D
2.6” H x 1.88” W x 1.1” D
Aerogen Pro-X Controller
Dimensions
33mm H x 75mm W x 131mm D
1.3” H x 2.9” W x 5.2D
Controller Cable Length 1.8 m (5.9 ft.)
AC/DC Adapter Cable Length 2.1 m (6.7 ft.)
Nebuliser Weight 13.5 g (0.5 oz) nebuliser and plug
Aerogen Pro-X Controller Weight 230 g (8.1 oz.), including battery and cable
Nebuliser Capacity Maximum 6 mL
Table 4. Environmental Specifications of the Aerogen Solo System
Operating
Maintains specified performance at circuit pressures up to 90cm
H
2
O and temperatures from 5 ºC (4F) up to 45 ºC (113ºF).
Atmospheric Pressure 450 to 1100 mbars
Humidity 15% to 95% relative humidity
Noise Level < 35 dB measured at 0.3 m distance
Storage & Transport
Transient Temperature
Range
-20 to +6C (-4 to +140°F)
Atmospheric Pressure 450 to 1100 mbars
Humidity 15 to 95% relative humidity
Table 5. Power Specifications of the Aerogen Solo System
Power Source
FRIWO (AG-AP1040-XX*) AC/DC adapter (input 100 to 240 VAC
50 – 60 Hz, output 9 V) or internal rechargeable battery (4.8 V
nominal output).
Note: The Aerogen Pro-X Controller is approved for use
with Aerogen AC/DC adapter AG-AP1040-XX* (Manufacturer
Reference: FRIWO FW7660/09)
Power Consumption < 8 Watts (charging), 2.0 Watts (nebulising).
Patient Isolation
Controller circuitry provides 4 kilovolt (kV) patient isolation and
complies with IEC/EN 60601-1.
* Consult your local representative for the order number extension specific
to your country and for pricing information.