System Warnings
Read and study all instructions before using the Aerogen Solo
System and accessories. Only trained medical personnel should
operate the device.
This is a single patient use device not to be used on more than one patient
to prevent cross infection.
The components and accessories of the Aerogen Solo System, as
packaged, are not sterile.
The components and accessories of the Aerogen Solo System are not
made with natural rubber latex.
Inspect all parts before use, and do not use if any parts are missing,
cracked or damaged. In case of missing parts, malfunction or damage,
contact your sales representative.
Only use physician-prescribed solutions that are approved for use with
a general purpose nebuliser. Consult drug manufacturer’s instructions
regarding suitability for nebulisation.
Use only with Aerogen Solo components, connectors and any accessories,
which are specified by Aerogen in this instruction manual.
Do not use beyond defined life (see page 25).
Do not use in the presence of flammable substances or flammable
anaesthetic mixtures combined with air, oxygen or nitrous oxide.
To avoid the risk of fire, do not use to aerosolise alcohol-based medications,
which can ignite in oxygen-enriched air under high pressure.
Do not autoclave any component or accessory of the Aerogen Solo
System.