EasyManua.ls Logo

Aerogen Pro-X - Nebulisation Modes; 30 Minute Mode (Intermittent)

Aerogen Pro-X
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Aerogen
®
Solo System Instruction Manual
17
Nebulisation Modes
30 Minute Mode (Intermittent)
Warnings
To avoid damage to the Aerogen Solo, do not use a syringe with a
needle.
During use observe for correct functioning of the nebuliser.
The maximum capacity of the nebuliser is 6 mL.
For intermittent doses less than or equal to 6 mL:
1. Open the plug on the nebuliser.
2. Use a pre-filled ampoule or syringe to add medication into the filler port
of the nebuliser (Figure 12).
3. Close the plug.
Figure 12. Filling the nebuliser with a pre-filled ampoule
4. To start a 30 Minute nebulisation cycle, press and release the blue
On/Off power button (Figure 4). The green 30 Minute indicator light
illuminates to indicate that the 30 Minute nebulisation cycle is in
progress.
5. To stop the nebuliser at any time, press the On/Off power button. The
indicator turns off to indicate that nebulisation has stopped.
Note: Medication can be added to the Aerogen Solo during nebulisation.
This does not interrupt nebulisation or ventilation.