Aerogen
®
Solo System Instruction Manual
31
Appendix 1: EMC Tables
The following tables are provided in accordance with IEC/EN 60601-1-2:
Guidance and manufacturers declaration electromagnetic emissions
The Aerogen Solo nebuliser system is intended for use in the electromagnetic
environment specified below. The customer or the user of the Aerogen Solo nebuliser
system should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic Environment -
Guidance
RF Emissions
CISPR 11
Group 1 The Aerogen Solo nebuliser system
uses RF energy only for its internal
function. Therefore, its RF emissions
are very low and are not likely to cause
any interference in nearby electronic
equipment.
RF Emissions
CISPR 11
Class B The Aerogen Solo nebuliser system is
suitable for use in all establishments,
including domestic establishments and
those directly connected to the public
low voltage power supply network that
supplies buildings used for domestic
purposes.
Harmonic emissions
IEC/EN 61000-3-2
Compliant
Voltage fluctuations /
flicker emissions
IEC/EN 61000-3-3
Compliant