Aerogen
®
16
1. Insert Aerogen Solo nebuliser firmly into Aerogen Ultra in orientation
shown in Figure 11.
2. If supplemental oxygen is required, firmly attach oxygen tubing to
Aerogen Ultra.
Note: Oxygen flow rate should be set between 1-6 LPM.
3. If a face mask is required, remove mouthpiece and attach face mask
to Aerogen Ultra.
Note: When using an open face mask, a minimum oxygen flow of 1 LPM
is required.
4. Add medication to nebuliser.
5. Connect cable to Aerogen Solo and power on controller.
6. Introduce Aerogen Ultra to patient and observe aerosol flow to ensure
correct operation.
7. Remove excess rainout from the Aerogen Ultra periodically (hourly with
continuous nebulisation).
8. To ensure optimum performance of the Aerogen Ultra, remove any
residue by rinsing through with sterile water, shake off excess and allow
to air dry.
Warnings
Do not use with a closed face mask.
When using with an open face mask, always use supplemental oxygen
flow of 1-6 LPM.
Performance of the Aerogen Ultra may vary depending upon the type
of drug and Aerogen Ultra configuration used.
Do not exceed recommended oxygen flow for system.
Ensure oxygen connection port or tubing is not occluded.
Do not use Aerogen Ultra without a mouthpiece or face mask.
Visually check Aerogen Ultra post-rinsing to ensure that valves have
not become dislodged.
Do not cover Aerogen Ultra valves during use.
Do not use Aerogen Ultra in conjunction with the Aerogen Pro.
Do not autoclave any component of the kit.
Ensure tubing is safely orientated to prevent strangulation hazard.