5
Do not modify this equipment without the authorisation of the manufacturer.
Do not use or store outside of specified environmental conditions.
To avoid damage to the nebuliser:
• Do not apply undue pressure to the domed aperture plate in the centre
of the nebuliser.
• Do not push out the Aerogen Vibronic
®
aerosol generator.
• Do not use a syringe with a needle to add medication.
• Do not attempt to clean the nebuliser.
Use of the Aerogen Solo and T-piece during the administration of volatile
anaesthetics may result in adverse effects on the constituent plastics. Do
not use with volatile anaesthetics unless known to be compatible. Aerogen
have determined that, using anaesthetic ventilators, the following volatile
anaesthetic agents are compatible under the stated conditions below:
Anaesthetic Agent Proprietary Name
Maximum
Percentage of
Anaesthetic
Maximum Duration
of Exposure
Isoflurane FORANE
®
3.5 % 12 hours
Sevoflurane SEVOFLURANE
®
8 % 12 hours
Desflurane SUPRANE
®
10 % 12 hours