Table 2. Aerogen Pro-X Controller Troubleshooting (Continued)
If this happens: It could mean: Try this:
Medication is left in
the nebuliser after
nebulisation cycle.
Nebuliser was not turned
on or connected to power.
Ensure that nebuliser is
connected to power and
turned on.
Rechargeable battery is
depleted.
Recharge battery (see
Recharging the Battery).
A 30 Minute cycle was
selected when connected
to the continuous feed
system.
Run a continuous cycle.
It may be time to replace
the nebuliser.
See Warranty and Life of
Product. Refer to Aerogen
Solo parts list by visiting
www.aerogen.com.
Flashing amber light.
It may mean that it is time
to replace controller.
Contact your local Aerogen
sales representative.
Note: The rechargeable battery in the Aerogen Pro-X Controller should
only be replaced by Aerogen authorised personnel: contact your Aerogen
sales representative.