Aerogen
®
12
2. The Aerogen Solo can be placed on the dry side of the humidifier as
shown in Figure 7. The Aerogen Solo can be used with a nasal interface
in this configuration.
Figure 7. Aerogen Solo on dry side of humidifier
3. Follow ventilator manufacturer instructions for performing a leak test
after inserting or removing the nebuliser.
Warnings
Condensate can collect and occlude ventilator circuits. Always position
ventilator circuits so that fluid condensate drains away from the patient.
Always connect a bacteria filter to the expiratory inlet of the ventilator.
Otherwise the function of the expiratory channel may be degraded.
Do not use a filter or heat-moisture exchanger (HME) between the
nebuliser and patient airway.