DD+DIS013.02E
Maintenance
Edition 1, Revision 10 ADC Solo Section 12 / 7
(Type 5155 / 100/200)
(3) Check roller A for visible wear
and replace if necessary.
(4) To replace the roller, remove
the retaining ring B and pull
the roller off its shaft.
In case of recorded cassette
feeding problems replace the
roller anyway and in any case
replace it once a year.
B
Figure 4
Belt
(5) Exchange the transport belt
once a year.
3.5 VME Rack
• Vacuum cleaner
VME fan
General
(1) Check the fan function of the
VME rack fan by hand.
There is no air filter to be
replaced.
(2) Remove visible dust and dirt
deposits by means of a
vacuum cleaner.