4 Checking the Image Quality
Check the last 20 to 40 images on the VIPS, to see if artifacts or other
image quality problems occur.
4.1 Test Cycles
(1) Carry out four test cycles with each format of the cassettes.
4.2 Exposure of a Flatfield
Expose an image plate of every format and evaluate all the images on the
Processing Station and a printer. Following, check the flat field for
homogenous field or stripes criteria. The hard disk of the digitizer provides two
flat field samples for quality comparison.
Repeat this procedure for all formats on site!
Flatfield
(1) Print the flat field sample provided by the digitizer:
Start the service program.
Select from the service menu
<Checks>
<Send flatfield>
<Calibration pattern>
<Banding pattern>
Print the flat fields "Calibration" and "Banding" via the Processing Station
(window setting of 0.6, without changing the level setting).