4 Remove the cassette from the cassette slot.
Reading an image plate in the dedicated
configuration
In the dedicated configuration, the CR 35-X can digitize the image plate while
you are entering the identification data via the ID Station.
To read an image plate:
1 Check that the CR 35-X is ready for operation:
• The CR 35-X must display the operator main screen with ‘Ready’ status, e.g.:
• The status indicator at the top of the CR 35-X must be green and be lit constantly.
2 Press the appropriate emergency button at the front of the CR 35-X.
The emergency button determines the speed class, i.e. the sensitivity, which will be
used to digitize the image plate.
WARNING: When the CR 35-X returns the cassette, it is ready to be re-
used immediately. However, if you leave it for more than 3 days before
re-using it, you must re-erase it first. Refer to ‘Re-erasing an image plate’
on page 56.
Note: If the CR 35-X displays an error message, refer to ‘Troubleshooting
checklist’ on page 65.
Note: The CR 35-X is operational if the status field equals ‘READY’, even if
status messages of the destination are shown (e.g. ‘QS not ready’).
QS not ready READY