66
4454C EN 2013-05-14Troubleshooting checklist
CR 35-X DIGITIZER
Status field:
Error field:
ERROR
‘CODE XXXXX
’
MESSAGE 1 Message 2 Action
POWER SUPPLY OUT OF
TOLERANCE
1. Check setting of
voltage selector switch
on back panel.
2. Check fuses of the
machine
3. Check supply
voltage.
Refer to ‘Checking the voltage supply’
on page 116 of the Reference manual.
IP JAM
1. Remove right side
panel
2. Put plate back into
cassette.
3. Close right side panel.
Refer to ‘Solving image plate and
cassette jams’ on page 119 of the
Reference manual.
Status field: WARNING
MESSAGE 1 Message 2 Action
ERASURE LAMP
[X], [Y], [Z] DEFECTIVE
Press to complete, IP
is not erased
• Press Confirm key.
• Refer to ‘Replacing the erasure
lamps’ on page 111 of the
Reference manual.
SCANNER WARNING
Possible bad image,
press
• Press Confirm key.
• Contact your local service
organization.
[PPNAME] NOT READY
Please check and press
• Check image processing station.
• If image processing station is
ready, press Confirm key.
CORRUPTED MESSAGE IN
QUEUE
Please press
and
check queue
• Press Confirm key.
• Refer to ‘Consulting the image
transmission queue (‘Queue
management’)’ on page 66 of the
Reference manual.
UNKNOWN DESTINATION
[PPNAME]
Please press
and
check queue
• Press Confirm key.
• Refer to ‘Consulting the image
transmission queue (‘Queue
management’)’ on page 66 of the
Reference manual.
ERROR WHILE LOADING
LANGUAGE FILE
Default language is
used, please press
• Press Confirm key. English will be
used.
• Restart CR 35-X.
• If the problem persists, contact
your local service organization.