4454C EN 2013-05-14 Specifications
CR 35-X DIGITIZER
Power consumption
Standby
230 V/ 50 Hz configuration 250 W
USA: 120 V/ 60 Hz configuration 250 W
Japan: 100 V/50 Hz and 100 V/60 Hz 250 W
During operation
230 V/ 50 Hz configuration max. 1610 W
USA: 120 V/ 60 Hz configuration max. 1440 W
Japan: 100 V/50 Hz and 100 V/60 Hz max. 1500 W
Environmental requirements
Room temperature 15 °C - 30 °C
Maximum temperature change 0.5 °C/min.
Relative humidity 15 % - 80 %
Magnetic field (Dynamic) compliant with EN 61000-4-8, Level 3
Sunlight exposure not be operated in full sunlight
Barometric pressure during operation 70 kPa to 106 kPa
Related altitude on site 3000 m to 0 m
Environmental conditions (during storage)
Room temperature between -25 °C and 55 °C
Relative humidity between 10 % and 100 %
Rate of change of temperature 0.5 °C/min
Barometric pressure between 70 kPa and 106 kPa
Environmental conditions (during transport)
Room temperature between -25 °C and 60 °C
Relative humidity between 10 % and 80 %
Barometric pressure between 70 kPa and 106 kPa