74
4454C EN 2013-05-14Specifications
CR 35-X DIGITIZER
Warming-up time
Cold start fully operational after max. 30 min.
Warm start
fully operational after self-test
if not switched off for more than 3 min., after 30
min. operation
Physical emissions
Noise emission (sound power level according to ISO 7779)
During scanning max. 65 dB(A)
•Standby max.45dB(A)
Radio frequency emission
according to EN 55011, Class B and Radio
frequency emission FCC, Part 15, Subchapter B,
Class B
Heat emission
During scanning max. 1610 W
Standby 250 W
Cassette return time
51 - 95 secs
Performance
Throughput 71 plates/h (size dependent)
End of Life
Estimated product life
(if regularly serviced and maintained according
to Agfa instructions)
7 yrs.
Preventive maintenance
Preventive maintenance frequency
Note: Preventive maintenance needs to be done
by an Agfa certified field service engineer.
Once a year or after 12000 cycles (whatever
comes first).
Grayscale resolution
Data acquisition 12 bits/pixel
Output to processor 12 bits/pixel