Electromagnetic emissions
This DR Detector has been tested for a normal hospital environment as
described below.
The user of the DR Detector should ensure that it is used in such an
environment.
Nevertheless the HF-emission and immunity can be influenced by connected
data cables depending on length and the manner of installation.
Emissions test
Compliance Electromagnetic Environment
Guidelines
RF emissions in accord-
ance with CISPR 11
Group 1 This DR Detector uses RF energy
for data transmission. Therefore,
its RF emissions can cause inter-
ferences in nearby electronic
equipment.
RF emissions in accord-
ance with CISPR 11
Class B The DR Detector is suitable for
use in all establishments, includ-
ing domestic establishments and
those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.
Harmonic emissions in ac-
cordance with IEC
61000-3-2
Complies
Class B
Voltage fluctuations /
flicker emissions in ac-
cordance with
IEC 61000-3-3
Complies
112 | DR 14s | Remarks for HF-emission and immunity