AGFA DR 14s - Artifact in DR Detector Images

AGFA DR 14s
118 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Artifact in DR Detector images
Details An artifact is visible in the images produced by a DR
Detector.
Cause Exposure conditions have changed significantly since
latest calibration.
Brief Solution Perform calibration of the DR Detector.
For details, refer to the DX-D DR Detector Calibration
Key User Manual (0134).