If the DR Detector has been dropped:
1. Visually check the DR Detector for deformations.
2. Perform a calibration of the DR Detector. For instructions, refer
to the DX-D DR Detector Calibration Key User Manual
(document 0134).
3. Perform a flat field exposure and check the image for visible
artifacts. Typical flat field exposure settings are 75 kV, 10 µGy,
large focus and using 1.5 mm Cu filter without grid.
CAUTION:
Damaged grid. Reduced image quality. Please handle the grids
with special care.
Topics:
DR Detector Battery
Safety directions for the power supply
DR 14s | Introduction to the DR Detector | 61