Note: Make sure that the battery is securely attached.
The detector starts up.
3. If the detector was turned off without removing the battery, use the power
button to turn on the detector.
Press and hold the power button for 1 second.
Figure 16: Power button
During startup the status indicator is flashing orange. After startup the
status indicator is lit green, indicating the power status.
4. Check the DR Detector status icon on the DR Detector Switch. If the
displayed status is error and the detector is shared between NX
workstations, it may still be connected to another NX workstation. Hold
the detector with its IR data port close to the IR Data Communication Unit
that is connected to the NX workstation.
All status indicators on the DR Detector are lit green. The DR Detector is
ready.
Before exposure make sure to check the equipment daily and confirm that it
works properly.
Related Links
DR 14s on page 18
Charging a battery on page 86
Detector Status Indicators on page 85
Registering the DR Detector to an NX Workstation using automatic registration
on page 93
Problem solving on page 98
68 | DR 14s | Getting started