72 400/54 Premium Shielded NMR Magnet System Product Overview
5 Agilent E5025 Cryogen Monitor
Technical specification
Helium probes supported Dual, standard type, max length 1.25m with
121 /m wire.
Nitrogen probes supported
400 to 800 MHz Agilent NMR magnets
User data display 8-digit red LED
User status display All status information is communicated via the
8-digit LED display.
User controls three push buttons: +,
, and -
Configuration / calibration of
Nitrogen and Helium probes
Can be performed via front panel. A portable
computer with an Ethernet connection can also
be used.
Default calibration For 400/54/ASP Premium Shielded magnet
Data logging
Cryogen levels are logged in the unit after
every read operation, and can be downloaded
to computer via RS-232 or Ethernet.
Estimated endurance: nine months.
Level read interval
Nitrogen levels are refreshed every cycle (~5
secs) and recorded every eight hours (or
manual reading) coincident with the helium
readings.
He level levels are recorded every 8 hours
during normal operation and every 30 seconds
during fill mode.
Nitrogen refill volume display
resolution
0% to 100% in 5% steps.
0% = bottom of refill volume
F = 100% or full
Display reading level flashes for levels ~ 5%
Helium refill volume display
resolution
0% to 100% in 1% steps
0% = bottom of refill volume
Display reading level flashes for levels ~ 0%
Absolute helium physical
resolution (that is, resolution
that the unit can measure)
< ±2mm
Head unit size 76mm wide, 44mm tall, 125mm long
Head unit mass Approx. 250g
Primary PSU size 50mm wide, 40mm tall, 95mm long
Power requirements
15 Watts power consumption
Universal mains input (100-240V 50/60Hz)
Connector A Underside, central, TNC for connection to
nitrogen level probe, and for unit mounting
Connector B Underside, far end, transverse, 15 D type, for
power and helium level probe
Connector C RJ45, back panel, for Ethernet communication