92 400/54 Premium Shielded NMR Magnet System Product Overview
6 Technical Specifications
The Cryostat
General Description
The cryostat is of conventional design, consisting of a central
all- welded stainless- steel helium vessel, surrounded by a
gas- cooled radiation shield and liquid nitrogen reservoir. The
complete assembly is contained in a stainless- steel outer
vacuum vessel with two service turrets located on top of the
cryostat. The turrets provide access to the helium reservoir
for the demountable magnet leads, helium level probe, and
helium transfer siphon. The outer vessel has a bottom- flange
closure constructed from aluminium, which is sealed to the
main body, and bore- tube by compressed o- ring seals. The
system is designed to be shipped without disassembly to
facilitate quicker and easier installation.
Specifications
The cryostat is generally as shown in drawing CNZ331842,
Issue F.
System finish
Dimensions
See customer interface drawings.
Table 13 System finish
Cryostat Light gray (RAL 7047), semi-gloss.
Stand and T-plate (where supplied) Charcoal gray (RAL 7011)
semi-gloss.
Table 14 Cryostat dimensions
Minimum ceiling height
with low ceiling installation equipment
2.94m
2.61m
System assembly Lift from below
Weight, magnet and cryostat 450 kg
Weight, operational including legs and
cryogens
630 kg

Table of Contents