Agilent E5025 Cryogen Monitor 5
400/54 Premium Shielded NMR Magnet System Product Overview 75
The Secondary Level SET-UP Menu
This secondary level menu changes the parameters and
set- up of the unit. These values are preset at the factory for
the magnet type before shipment.
Entering the secondary level SET-UP menu
1 From the cryogen display mode, press the and
buttons simultaneously. The unit displays “SEt UP ?”
2 Press and a confirmation “SurE ?” is displayed.
3 Press to enter the menu.
4 To scroll down the menu, press the button.
Scrolling down through the menu eventually returns to
the cryogen display mode.
Functions available on the secondary level SET-UP menu
This section describes each item in the secondary level
SET- UP menu in the order they appear on the display.
To change a value for any function:
1 Press to select the function when it appears on the
display.
2 When the current value is displayed, press the and
buttons as necessary to select the desired value.
3 Press to save the new value and exit to the Cryogen
Display mode.
4 You may be presented with a prompt to confirm the
action before the unit returns to the cryogen display
mode.
If the display is allowed to auto time- out back to the
cryogen display mode (after approximately 15 seconds)
without saving, then the value is unchanged and any actions
are ignored.
During normal operation, there is no need to access the secondary
level SET-UP menu other than to perform a recalibration of the
nitrogen probe.
Incorrect use of the SET-UP menu can damage the unit.