Agilent E5025 Cryogen Monitor 5
400/54 Premium Shielded NMR Magnet System Product Overview 77
diSP P1 ?
Selects P1 to be displayed in the cryogen display mode. This
option is only available when both the probes are enabled.
diSP P2 ?
Selects P2 to be displayed in the cryogen display mode. This
option is only available when both the probes are enabled.
n2 on ?
Enables the nitrogen level probe.
In order for the nitrogen to read correctly once enabled,
perform a calibration while the nitrogen vessel is full (see
CAL n2 ? below).
n2 oFF ?
Disables the nitrogen level probe. The nitrogen reading is
displayed as “oFF”.
CAL n2 ?
Performs a calibration of the nitrogen probe. Make sure that
the nitrogen vessel is FULL when the routine is run.
Otherwise it will calibrate incorrectly. The unit displays
CAL n2 - ” during calibration, then returns to Cryogen
Display mode.
FAC CAL?
Recalls the factory saved calibration values. It does not reset
the P1 or P2 display settings.
The unit displays “rECAL - ” during calibration and confirms
donE” when completed, then returns to Cryogen Display
mode.
iP 1 0 ?
Sets the IP address of the unit to: IP address [10.0.0.241],
Gateway [10.0.0.1], Subnet mask [255.255.255.0].
iP 1 72 ?
Sets the IP address of the unit to: IP address [172.16.0.241],
Gateway [172.16.0.1], Subnet mask [255.255.255.0].