EasyManua.ls Logo

Agilent Technologies 5977 MSD

Agilent Technologies 5977 MSD
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installing GC Columns 2
5977B Series MSD Operation Manual 35
7 Position the column so it extends 4 to 6 mm past the end of the ferrule
(Figure 4).
8 Slide the septum up to the bottom of the nut to fix the correct column
insertion length.
9 Insert the column in the inlet.
10 Slide the nut up the column to the inlet base and finger-tighten the nut.
11 Adjust the column position so the septum is even with the bottom of the
column nut.
12 Tighten the column nut an additional 1/4 to 1/2 turn. The column should
not slide with a gentle tug.
13 Start carrier gas flow.
14 Verify flow by submerging the free end of the column in isopropanol. Look
for bubbles.
See also
For more information about installing a capillary column, refer to Optimizing
Splitless Injections on Your GC for High Performance MS Analysis, Agilent
Technologies publication number 5988-9944EN.
Figure 4 Installing a capillary column for a split/splitless inlet
Insulation cup
Reducing nut
Capillary column
Ferrule (inside nut)
Inlet column nut
Septum
4 to 6 mm

Table of Contents

Related product manuals