4Operating in EI Mode
84 5977B Series MSD Operating Manual
To enable/disable BOOTP on MSD
By default, the BOOTP is disabled. If your LAN uses a BootP server, Enabling
BOOTP causes the server to automatically assign an IP address to the MSD.
1 Press Config, and then press MS/Aux Det to bring up the CONFIGURE MS
DETECTOR menu.
2 Press the down arrow to scroll to MSD BOOTP.
3 To enable BOOTP, press On/Yes.
To disable BOOTP, press Off/No.
4 Wait for the MSD to confirm the change on the GC control panel.
5 Reboot the MSD. (See “To reboot the MSD from the GC control panel” on
page 83.)