Exercise 1 – Develop an acquisition method
Task 5. Find optimum collision energy for MRM acquisition
32 Agilent 6400 Series Triple Quad LC/MS Familiarization Guide
h Click the Overlaid Mode icon,  .
i Compare peak intensities for each 
compound transition in each data file 
in the Chromatogram Results window.
• Compare the colors shown in 
Chromatogram Results with the 
color next to the MRM transition 
name in the Data Navigator.
• You can also right-click the 
Chromatogram Results window 
header and compare the colors of 
the chromatograms to the colors of 
the titles in the shortcut menu.
Unless you decide to acquire MRMs at 
lower collision energies, you should 
find that the optimal method settings 
are as shown in Table 4.
j Click the Close Data File icon in the 
main toolbar, and click Close when the 
Close Data File dialog box appears.
• You now have all the information 
you need to do an MRM acquisition 
experiment of the sulfa drug 
mixture. Consider doing at least 
one more run with those settings.
Steps Detailed Instructions Comments
Table 4 Compounds and Collision Energy
Compounds MRM Transition Collision Energy (V)
Sulfamethizole 271.0 > 155.8 10
Sulfamethazine 279.0 > 185.7 15
Sulfachloropyridazine 285.0 > 155.7 10
Sulfadimethoxine 311.0 > 155.7 15