Exercise 3 – Create a Triggered Dynamic MRM acquisition method
Task 2. Add/Modify compounds in an existing database
Agilent 6400 Series Triple Quad LC/MS Familiarization Guide 49
11 In the Database Browser program,
select the transitions.
a Mark the Show All Records check
box.
b Click the Select top button under
Select Transitions.
c Type
10 for the ranked transitions.
d Click the Select Transitions button.
All the transitions that you typed in
are visible.
The tools to allow you to set up
Primary transitions and Secondary
transitions are available in this
program.
If any of the compounds have two
collision energies, clear the check
boxes for one of these values.
See Table 5 on page 54 for values to
use in the method.