Exercise 3 – Create a Triggered Dynamic MRM acquisition method
Task 2. Add/Modify compounds in an existing database
50 Agilent 6400 Series Triple Quad LC/MS Familiarization Guide
12 In the Database Browser program,
automatically select the Primary
transitions and Trigger transition.
a In the Set top ranked transitions as
primary box, enter 2.
b Click the Set Primaries and Trigger
button.
The program automatically selects
the two most abundant transitions
as the Primary transitions.
The program also selects the most
abundant transition as the Trigger.
You can manually select a second
Trigger transition.
See Table 5 on page 54 for values to
use in the method.
Steps Detailed Instructions Comments
You examine the Primary column and the Trigger column to determine which transitions are selected. You can select
one or two Trigger transitions. You can select multiple Primary transitions.