26 Advanced Operation
2 How the 7697A Headspace Sampler Works
How the HS Reduces Carryover
The 7697A HS provides two special features to reduce
carryover.
After each vial, the HS purges the sample loop and
sample probe with a high flow of vial pressurization gas,
as defined in the method. This is called the Purge flow,
and you control both the flow rate and purge time.
Between each sequence, the HS purges the sample loop
and sample probe with a continuous, low flow of vial
pressurization gas. This is called the Standby flow. You can
control the flow rate.