EasyManuals Logo

Agilent Technologies 7697A Advanced Operation Guide

Agilent Technologies 7697A
90 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #26 background imageLoading...
Page #26 background image
26 Advanced Operation
2 How the 7697A Headspace Sampler Works
How the HS Reduces Carryover
The 7697A HS provides two special features to reduce
carryover.
After each vial, the HS purges the sample loop and
sample probe with a high flow of vial pressurization gas,
as defined in the method. This is called the Purge flow,
and you control both the flow rate and purge time.
Between each sequence, the HS purges the sample loop
and sample probe with a continuous, low flow of vial
pressurization gas. This is called the Standby flow. You can
control the flow rate.

Table of Contents

Other manuals for Agilent Technologies 7697A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7697A and is the answer not in the manual?

Agilent Technologies 7697A Specifications

General IconGeneral
BrandAgilent Technologies
Model7697A
CategoryMedical Equipment
LanguageEnglish

Related product manuals