Early Maintenance Feedback 7
Advanced Operation 87
To Set Up EMF for Use
To use EMF:
1 After performing maintenance on an item (replacement or
calibration, for example), set a limit for its counter. See
To Set a Limit for a Counter.
2 The counter will increment as the HS performs injections
and runs sequences.
3 When the counter exceeds the limit, the HS will light the
Service Due indicator and log an entry into the Event log.
4 Replace or calibrate the item.
5 Reset the item’s counter. See To Reset a Counter.