452 Chapter7
Language Reference
DEMODT Demodulation Time
Query Response
Figure 7-92 DEMODT Query Response
Example
10 OUTPUT 718;"IP;"
20 OUTPUT 718;"FA 88MHZ;FB 108MHZ;"
30 OUTPUT 718;"MKN EP;"
40 PRINT "MOVE MARKER TO SIGNAL TO BE DEMODULATED; PRESS H
OLD."
50 PRINT "THEN PRESS CONTINUE"
60 PAUSE
70 INPUT "ENTER DEMODULATION TIME (.1 SEC - 60 SEC)",Dtime
80 OUTPUT 718;"DEMODT ";Dtime;"S;"
90 OUTPUT 718;"DEMOD FM;"
100 LOCAL 718
110 PRINT "ADJUST VOLUME AND SQUELCH AS NECESSARY."
120 END