Page #602 background image
602 Chapter7
Language Reference
SRCFINTK Fine Tracking Adjust
SRCFINTK Fine Tracking Adjust
Syntax
Figure 7-249 SRCFINTK Syntax
Description
The SRCFINTK command controls the fine adjustment of the frequency
of the built-in tracking-generator oscillator. Once enabled, this
adjustment is made in digital-to-analog-converter (DAC) values from 0
to 255. For coarse adjustment, refer to the SRCCRSTK command
description. SRCFINTK is available only with an 8560E/EC Option
002.
Parameters
number integer from 0 to 255.
UP or DN increments by 1.
Preset State
128