FULBAND Full Band
FULBAND Full Band
Syntax
Figure 7-121 FULBAND Syntax
NOTE When in preselected external mixing mode, band K is not available.
Description
The FULBAND command selects a commonly-used, external-mixer
frequency band, as shown in Table 7-1 on page 372. The harmonic lock
function (HNLOCK) is also set; this locks the harmonic of the chosen
band. External-mixing functions are not available with an 8560E/EC
Option 002.
Table 7-5 Unpreselected External-Mixer Frequency Bands
Frequency
Band
Frequency
Range(GHz)
Mixing
Harmonic
Conversion
Loss
K 18.0 to 26.5 6 30 dB
A 26.5 to 40.0 8 30 dB
Q 33.0 to 50.0 10 30 dB
U 40.0 to 60.0 10 30 dB
V 50.0 to 75.0 14 30 dB
E 60.0 to 90.0 16 30 dB
W 75.0 to 110.0 18 30 dB
F 90.0 to 140.0 24 30 dB