G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide 11
Reference: For 7700s
The recommended tuning parameter values for 7700s are shown below.
*1 Generally, set the QP Bias at least 2 to 3V higher than OctP Bias.
*2 High-Energy Collision mode allows for enhanced reduction of interference under the Helium mode.
However, as it will also result in reduced low-mass sensitivity, it should be used in accordance with
the purpose of the analysis.
Table 3 Recommended Values for 7700s (When Using a MicroFlow Nebulizer)
Parameter High Sensitivity Cell Gas Mode High Energy Collision
Mode
*2
Cool Plasma
Recom-
mended
Value
Recom-
mended
Range
Recom-
mended
Value
Recom-
mended
Range
Recom-
mended
Value
Recom-
mended
Range
Recom-
mended
Value
Recom-
mended
Range
RF Power [W] 1500 fixed 1500 fixed 1500 fixed 600 fixed
Smpl Depth [mm] 8 7 to 10 8 7 to 10 8 7 to 10 18 fixed
Carrier Gas [L/min] 0.7 fixed 0.7 fixed 0.7 fixed 0.7 fixed
Makeup Gas [L/min] 0.5 0.3 to 0.7 0.5 0.3 to 0.7 0.5 0.3 to 0.7 0.75 0.6 to 1.2
Dilution Gas [L/min] 0 fixed 0 fixed 0 fixed 0 fixed
Neb Pump [rps] 0.1 fixed 0.1 fixed 0.1 fixed 0.1 fixed
S/C Temp [degC] 2 fixed 2 fixed 2 fixed 2 fixed
He or H
2
gas [ml/min] 0 fixed 3.6(He)
4.0(H
2
)
3.2 to 4.0(He)
3.6 to 4.4(H
2
)
10(He) 7 to 10(He) 0 fixed
Extract 1 [V] 4.5 3 to 7 4.5 3 to 7 4.5 3 to 7 -120 -200 to -40
Extract 2 [V] -100 -170 to -60 -100 -170 to -60 -100 -170 to -60 -5 -30 to 5
Omega Bias [V] -70 -100 to -30 -70 -100 to -30 -70 -100 to -30 -70 -120 to -30
Omega Lens [V] 11 5 to 15 11 5 to 15 11 5 to 15 6 3 to 10
Cell Entrance [V] -30 -40 to -30 -40 -40 to -30 -130 -150 to -110 -30 -40 to -30
Cell Exit [V] -50 -60 to -40 -60 -60 to -40 -150 fixed -60 fixed
Deflect [V] 12 8 to 15 0 -5 to 4 -80 -90 to -70 9 5 to 13
Plate Bias [V] -40 -50 to -30 -60 fixed -150 fixed -60 fixed
OctP RF [V] 180 150 to 200 180 150 to 200 -190 180 to 200 150 100 to 200
OctP Bias [V]
*1
-8 -12 to -6 -18 fixed -100 fixed -18 -30 to -10
QP Bias [V]
*1
-3 -5 to -3 -15 fixed -96 -97 to -90 -5 -5 to -3