22 G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
3 Fill in values for the rest of your samples as follows:
a Select the first row and the subsequent rows by dragging the mouse.
b Right- click in the highlighted rows and select Fill Down from the
shortcut menu.
c The values for Method, Type, Dil/LVL and ISTD Conc are copied from
Row 1 to the selected blank rows. Sequential Vial numbers are assigned.
4 Modify the information for the sample rows added in Step 3 to match your
analyses.
5 (optional) To turn off the plasma automatically at the end of the sequence,
append the following row to the end of the sequence:
• Type: Keyword
• Data File: StandBy
• Leave all other cells blank
6 (optional) Select the Import DA Method from Existing Batch check box to
import an existing data analysis method from the specified data batch
directory. You can import the DA Method Only or the DA method and Std
Data. You can also apply a data analysis method after data acquisition.
7 Click OK to close the Edit Sample Log Table dialog box.
8 If errors are found in the table, a message appears.
a Click OK and the offending cell will be highlighted in the table.
b Fix the problem in the highlighted cell.
9 Repeat Steps 7 & 8 until no errors are detected.
10 (optional) Select Save from the Sequence menu in the ICP- MS Top window.
Enter a name for the sequence you just created, and click OK. You don’t
need to save a sequence to run it, but doing so allows you to open it later
and use it as a starting point for creating a new sequence more quickly.