11
Stopping or pausing a run
Bravo Platform Quick Guide
Stopping or pausing a run
Emergency stops
An emergency stop might be necessary to prevent a collision of the Bravo head with
another object, such as misaligned labware. If you want to temporarily pause and then
continue a run, see “Pausing and continuing a run” on page 12.
To stop in an emergency:
Press the red button on the emergency-stop pendant. The Bravo head stops moving.
Note: Interrupting the light curtain also stops the motion of the Bravo head.
Figure
Emergency-stop pendant
A Bravo Error dialog box opens and displays a robot disable is active error message.
To restore the Bravo device after an emergency stop:
1 At the emergency-stop pendant, turn the red button clockwise. The spring-loaded
button pops up.
2 Ensure that no object is interrupting the Light Curtain.
3 In the Bravo Error dialog box, click Retry. The Bravo Platform will attempt to resume
the run where it left off.
To abort the run after an emergency stop:
1 In the Bravo Error dialog box, click Abort.
2 In the Scheduler Paused dialog box, click Abort process.

Other manuals for Agilent Technologies Bravo Platform

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Bravo Platform and is the answer not in the manual?

Agilent Technologies Bravo Platform Specifications

General IconGeneral
BrandAgilent Technologies
ModelBravo Platform
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals