Introduction
14 Cary 50 Spectrophotometer Hardware Manual
The power requirements of the Cary 50 must be met by the computer
power supply and the following capacity (in addition to what is
required by the computer itself) is required:
+5 V DC <1 A
+12 V DC <1.5 A
-12 V DC <0.25 A
a total of 26 W for the instrument
Allowance must also be made for external accessories powered by
+12 V DC.
Operating motor-driven accessories will increase the +12 V current
by a further 2 A (a maximum of 24 W for accessories).
Agilent recommends a computer power supply rating of 220 W as a
minimum.
Connections
3.5 mm phono jack socket in the left side of the sample
compartment for accessories.
8-pin DIN connector in the left side of the sample compartment
for the diode detector.
25-pin D-range connector in the right side of the sample
compartment for accessories (optional).