EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>Cary 50

Agilent Technologies Cary 50 User Manual

Agilent Technologies Cary 50
38 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
Maintenance
Cary 50 Spectrophotometer Hardware Manual 25
Aligning the source mirror
For optimum performance of the instrument, the source mirror
should be aligned to suit the new lamp module.
To align the source mirror:
1 Start the Align application by clicking the Windows Start button
and choosing (All) Programs, Agilent, Cary WinUV, (or Agilent,
UV/UV FO Dissolution) then Align.
2 On the Cary tabbed page, set the following parameters:
Beam Mode: Single Beam, Normal
Y Mode: %T
Ave. Time: 0.5000
Wavelength: 500 nm
3 Click Apply.
4 Remove the snap-out panel (refer to Figure 5) from the front of
the instrument, using a flat-blade screwdriver to pry the panel
open. This exposes the two source mirror adjustment screws.
NOTE

Other manuals for Agilent Technologies Cary 50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Cary 50 and is the answer not in the manual?

Agilent Technologies Cary 50 Specifications

General IconGeneral
BrandAgilent Technologies
ModelCary 50
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals