Troubleshooting
Cary 50 Spectrophotometer Hardware Manual 35
Wavelength calibration
Problem
The position of an absorption (transmission or reflectance) peak
appears to shift over a prolonged period of time.
Solution
Perform a wavelength calibration.
Restart the computer. Agilent recommends that as part of your
standard operating procedure, you restart the computer attached
to the Cary 50 monthly.
Perform the Instrument Performance Test suite in ‘Validate’ to
ensure the instrument results meet specification.