Troubleshooting
Cary 50 Spectrophotometer Hardware Manual 33
Figure 7. Wild absorbance fluctuation during a scan
Solution
The detector cable inside the Cary 50 sample compartment is not
connected. Check that the 8-pin plug on the left side of the sample
compartment is firmly plugged in.
Instrument performance testing
Problem
The results of your instrument performance tests do not meet
specifications (the results obtained during factory testing are
included in the packing crate with the instrument).
Solution
Check that:
The sample compartment is empty.
The cable connecting the instrument to the computer is correctly
connected and the retaining screws are tightened.