EasyManua.ls Logo

Agilent Technologies E3631A - Safety Considerations

Agilent Technologies E3631A
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 1 Specifications
Performance Specifications
17
1
Performance Specifications
Output Ratings (@ 0°C - 40°C)
+6V Output 0 to +6 V ; 0 to 5 A
+25V Output 0 to +25 V ; 0 to 1 A
-25V Output 0 to -25 V ; 0 to 1 A
Programming Accuracy
[1]
12 months (@ 25°C ± 5°C), ±(% of output + offset)
+6V Output
+25V Output -25V Output
Voltage 0.1% + 5 mV 0.05% + 20 mV 0.05% + 20 mV
Current 0.2% + 10 mA 0.15% + 4 mA 0.15% + 4 mA
Readback Accuracy
[1]
12 months (over GPIB and RS-232 or front panel with respect
to actual output @ 25°C ± 5°C), ±(% of output + offset)
+6V Output
+25V Output -25V Output
Voltage 0.1% + 5 mV 0.05% + 10 mV 0.05% + 10 mV
Current 0.2% + 10 mA 0.15% + 4 mA 0.15% + 4 mA
Ripple and Noise (with outputs ungrounded, or with either output terminal grounded,
20 Hz to 20 MHz)
+6V Output
+25V Output -25V Output
Voltage <0.35 mV rms <0.35 mV rms <0.35 mV rms
<2 mV p-p <2 mV p-p <2 mV p-p
Current <2 mA rms <500 μA rms <500 μA rms
Common mode current <1.5 μA rms
Load Regulation,
±(% of output + offset)
Change in output voltage or current for any load change within ratings
Voltage <0.01% + 2 mV
Current <0.01% + 250 μA
Line Regulation,
±(% of output + offset)
Change in output voltage and current for any line change within ratings
Voltage <0.01% + 2 mV
Current <0.01% + 250 μA
[1]
Accuracy specifications are after an 1-hour warm-up and calibration at 25°C

Table of Contents

Other manuals for Agilent Technologies E3631A

Related product manuals