EasyManuals Logo
Home>Agilent Technologies>Power Supply>E3647A

Agilent Technologies E3647A User Manual

Agilent Technologies E3647A
211 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #199 background imageLoading...
Page #199 background image
Appendix Service Information
General Calibration/Adjustment Procedure
199
Service
Information
5 Save the changes and select the middle voltage calibration point.
If the entered number is within an acceptable range, an ‘ENTERED’ message
appears for a second. If the entered number is not correct, an error message will be
displayed for a second and you will hear a beep, and then go back to the low, middle,
or high voltage calibration point again as proceeding.
6 Enter the reading you obtained from the DVM by using the knob and resolution
selection keys.
7 Save the changes and select the high voltage calibration point.
8 Enter the reading you obtained from the DVM by using the knob and resolution
selection keys.
9 Save the changes and go to the OVP calibration mode.
If the calibration fails, an error message will be displayed for a second and go back
to the voltage calibration mode again. A ‘VOLTAGE CAL’ message is displayed.
Above message is displayed to indicate that the power supply is ready for the OVP
calibration.
V MI 10.000
V MI 11.058
V HI 19.500
V HI 19.495
ovp1 cal
Calibrate
View
Calibrate
View
Calibrate
View

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3647A and is the answer not in the manual?

Agilent Technologies E3647A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3647A
CategoryPower Supply
LanguageEnglish

Related product manuals