Appendix Service Information
General Calibration/Adjustment Procedure
201
Service
Information
14 Enter the computed value (DVM reading ÷ by shunt resistance) by using the
knob and resolution selection keys.
15 Save the change and select the high current calibration point.
16 Enter the computed value (DVM reading
÷ by shunt resistance) by using the
knob and resolution selection keys.
17 Save the new current calibration constants for the output1 and go to the
calibration mode for the output2.
If the calibration fails, an error message will be displayed for one second and you will
hear a beep, and then go back to the current calibration mode again.
18 Repeat the steps (3) through (16) for the voltage, OVP, and current calibration.
19 Exit the calibration mode.
The above message appears if the calibration is successful. If the calibration fails, an
error message will be displayed for one second and you will hear a beep, and then go
back to the current calibration mode again.
I MI 1.5400
I HI 2.8000
I HI 2.789 A
volt2 cal
cal mode
Calibrate
View
Calibrate
View
Power

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3647A and is the answer not in the manual?

Agilent Technologies E3647A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3647A
CategoryPower Supply
LanguageEnglish

Related product manuals