3-8 Chapter3
Functional Testing
Displayed Average Noise Level
4. Press the following keys on the analyzer:
Sweep, Sweep (Cont)
BW/Avg, Average (Off)
BW/Avg, Resolution BW (Auto)
Video BW (Auto)
SPAN, 50 kHz
FREQUENCY
5. Press Center Freq. Set the center frequency of the analyzer to the
frequency recorded in the Measured Frequency column (b) in
Table 3-1 for 501 MHz to 1.0 GHz.
6. Press the following keys on the analyzer:
BW/Avg, 1 kHz
Video BW Auto Man, 30 Hz
Single
Wait for the sweep to finish.
7. Press the following keys on the analyzer:
Display, Display Line (On)
Adjust the display line so that it is centered on the average trace
noise, ignoring any residual responses.
8. Record the display line amplitude setting as TR Entry 2 (TR Entry
17 for a 75 Input) in Table 3-1. The Average Noise Level should be
less than the Maximum for the appropriate input impedance.