Chapter 3 3-17
Functional Testing
Displayed Average Noise Level
4. Press the following keys on the analyzer:
Sweep, Sweep (Cont)
BW/Avg, Average (Off)
BW/Avg, Resolution BW (Auto)
Video BW (Auto)
SPAN, 50 kHz
FREQUENCY
5. Press Center Freq. Set the center frequency of the analyzer to the
frequency recorded in the Measured Frequency column (i) in
Table 3-1 for 6.01 GHz to 12.0 GHz.
6. Press the following keys on the analyzer:
BW/Avg, 1 kHz
Video BW Auto Man, 30 Hz
Single
Wait for the sweep to finish.
7. Press the following keys on the analyzer:
Display, Display Line (On)
Adjust the display line so that it is centered on the average trace
noise, ignoring any residual responses.
8. Record the display line amplitude setting as TR Entry 12 in
Table 3-1. The average noise level should be less than the Maximum.
Procedure (12.01 GHz to 13.2 GHz) E4405B
1. Press Preset System, Power On/Preset, Preset (Factory), Preset, on the
analyzer. Wait for the preset routine to finish. Set the analyzer by
pressing the following keys:
FREQUENCY, Start Freq, 12.01 GHz
Stop Freq, 13.2 GHz
AMPLITUDE, –70 dBm
Attenuation (Man), 0 dB
BW/Avg, 1 MHz
Video BW Auto Man, 10 kHz

Table of Contents

Other manuals for Agilent Technologies E4403B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4403B and is the answer not in the manual?

Agilent Technologies E4403B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4403B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals