Chapter 3 3-25
Functional Testing
Frequency Readout Accuracy
5. Repeat step 3. If the instrument is functioning correctly, the result
should be between 1.4899990 GHz and 1.4900009 GHz.
NOTE The Frequency Readout Accuracy in now complete. Continue with the
Marker Count Accuracy functional check.