EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>E4418B

Agilent Technologies E4418B Service Guide

Agilent Technologies E4418B
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #26 background imageLoading...
Page #26 background image
Specifications
Power Meter Supplemental Characteristics
1-8 Agilent E4418B/E4419B Service Guide
Table 1-3: Power Sensor Specifications±
Power Sensor Zero Drift
1
1. Within 1 hour after zero set, at a constant temperature, after a 24 hour
warm-up of the power meter.
Measurement Noise
2
2. The number of averages at 16 (for normal mode) and 32 (for x2 mode), at
a constant temperature, measured over a 1 minute interval and 2 standard
deviations. For Agilent E-series power sensors the measurement noise is
measured within the low range. Refer to the relevant power sensor manual
for further information.
Agilent 8481A <±10 nW <110 nW
Agilent 8481B <±10 µW <110 µW
Agilent 8481D <±4pW <45 pW
Agilent 8481H <±1 µW <10µW
Agilent 8482A <±10 nW <110 nW
Agilent 8482B <±10 µW <110 µW
Agilent 8482H <±1 µW <10µW
Agilent 8483A <±10 nW <110 nW
Agilent 8485A <±10 nW <110 nW
Agilent 8485D <±4pW <45 pW
Agilent R8486A <±10 nW <110 nW
Agilent R8486D <±6pW <65 pW
Agilent Q8486A <±10 nW <110 nW
Agilent Q8486D <±6pW <65 pW
Agilent W8486A <±40 nW <450 nW
Agilent 8487A <±10 nW <110 nW
Agilent 8487D <±4pW <45 pW
Agilent E4412A 15 pW <70 pW
Agilent E4413A 15 pW <70 pW
4402serv.book Page 8 Tuesday, October 14, 2003 3:18 PM

Table of Contents

Other manuals for Agilent Technologies E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4418B and is the answer not in the manual?

Agilent Technologies E4418B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals