Performance Tests
Power Reference Level Test
2-16 Agilent E4418B/E4419B Service Guide
If an equivalent measurement system is employed for post-factory test,
the power reference oscillator output can again be set to 1 mW ±0.4%. This
setting is sufficient to ensure that the power reference oscillator meets the
accuracy specification of ±0.5% after 2 years. This specification includes
the ±0.4% setting and also takes into account other factors that will
contribute to determining the accuracy after 2 years. The complete
breakdown of the ±0.5% specification is as follows:
Test line limit: 0.1%.
System uncertainty: 0.3%.
2 year oscillator drift: 0.04%.
Delta environment: <0.06%.
The same equipment used to measure the accuracy of the power reference
oscillator can also be used to set it to the post-factory accuracy of ±0.4%
(for further information refer to chapter 3, Adjustments).
The following procedure ensures maximum accuracy in measuring power
reference oscillator output using Agilent test instruments of known
capability. Signal acquisition criteria may vary if alternative test
instruments are used. In such cases refer to the manufacturer’s guidelines
for further information.
The test is performed using Equation 2-4:
Equation 2-4: Power Level Provided By The Reference Oscillator
The components of this equation are defined as follows:
•V
0
is the voltage measured between V
comp
and V
RF
with no power
applied and after the Agilent 432A has been zeroed.
•V
1
is the voltage measured between V
comp
and V
RF
with power applied.
This power is approximately 1 mW.
•V
comp
is the voltage between the Agilent 432A V
comp
connector and
chassis ground with power applied.
R is the resistance of the mount resistor in the Agilent 432A power
meter. For further information refer to Appendix A.
Calibration Factor is the value of the thermistor mount at 50 MHz
P
meas
2V
comp
V
1
V
0
()V
0
2
V
1
2
+
4R Calibratio
nFactor()
-----------------------------------------------------------------------------------=
4402serv.book Page 16 Tuesday, October 14, 2003 3:18 PM

Table of Contents

Other manuals for Agilent Technologies E4418B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4418B and is the answer not in the manual?

Agilent Technologies E4418B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4418B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals