Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
183
Front-Panel Key and SCPI Command
Reference
4.4.3.1.3 Ratio Reference
If the Tuned RF Level is to be displayed relative to a reference, enter the value as a ratio reference using this key.
4.4.3.1.4 Ratio Mode
The ratio and previous ratio functions permit any measurement result to be scaled to a reference. The reference
may be either a measurement result or a keyboard entry.
The Log/Linear key enables measurement results to be displayed in linear or logarithmic units, both absolute and
relative. Examples of absolute units are kHz and mV (linear), or dBm (logarithmic). Examples of relative units
are % and dB. When the instrument displays dB as a relative unit, the “REL” indicator is shown on the display.
Example: DISP:TRFL:UNIT DBM
DISP:TRFL:UNIT?
Mode: MRECEIVE
Key Path:
Amplitude / Y Scale
Remote Command: :DISPlay:TRFLevel:RREFerence <real>
:DISPlay:TRFLevel:RREFerence?
:DISPlay:TRFLevel:RREFerence:AUTO 0|1|OFF|ON
:DISPlay:TRFLevel:RREFerence:AUTO?
Preset: 0dBm
State Saved: Saved in instrument state.
Unit: dBm
Min: –150dBm
Max: 30dBm
Example: DISP:TRFL:RREF 0.0
DISP:TRFL:RREF?
:DISP:TRFL:RREF:AUTO OFF
:DISP:TRFL:RREF:AUTO?
Mode: MRECEIVE
Key Path:
Display
Remote Command: :DISPlay:TRFLevel:RMODe LOG|LINear
:DISPlay:TRFLevel:RMODe?
Preset: LOG
State Saved: Saved in instrument state.
Range: Log | Linear